Photoionization Detectors (PID)

MiniRAE 3000 +

The most advanced wireless handheld VOC monitor with parts per million measurement

Overview
Whether for industrial hygiene, leak detection, or HazMat response, the MiniRAE 3000 + delivers one of the most advanced VOC monitoring capabilities on the market. With a photoionization detector (PID) that has an extended detection range from 0-15,000 ppm, a rapid three-second response time, built-in correction factors for more than 200 compounds, patented auto-cleaning sensor technology and Reflex PID Technology™, the MiniRAE 3000 + can measure one of the highest levels of ionizable chemicals available on the market.The powerful sample pump draws through up to 100 feet of tubing horizontally or vertically. A built-in flashlight to help navigate in darker environments and an IP 65/67 dust-tight and water-proof housing all add up to a device that performs reliably even in harsh conditions.MiniRAE 3000 + is solution oriented through multiple connected options. A dedicated mobile APP has been developed for Remediation engineers and Industrial Hygienists. Instrument is also available with real time connectivity which can relay readings and alarm status up to 2 miles to a RAELink 3 wireless transmitter.Features and Benefits:
  • Most advanced handheld VOC monitor on the market
  • PID measures from 0-15,000 ppm
  • 3-second response
  • Humidity compensation
  • Patented PID lamp auto-cleaning technology
  • Patented Reflex PID Technology™
  • Optional Mesh Radio for ConneXT compatibility
  • Optional built-in Bluetooth transmitter that transmits real-time data up to 2 miles with RAELink3 wireless router
  • Optional BLE Module
  • Sample pump draws from up to 100 feet
  • Correction factors for more than 200 compounds
  • Built-in flashlight
  • IP 65/67 water-proof and dust-tight
  • Logs up to 6 months of data at 1-minute intervals
  • Multi-language capability – up to 10 languages
  • 9.8eV for benzene and TVOC and 10.6 and 11.7 UV lamps for TVOC
Application:
  • Emergency Response
  • Environmental Cleanups/ Remediation
  • Hazmat Response
  • Leak Detection
  • Urban Search & Rescue
  • Venue Protection
  • Worker Exposure
Even the best products deserve expert maintenance!

Elevate your gas detectors’ potential with Honeywell's expert maintenance services! Maximize uptime, eliminate costs, and ensure on-time, on-budget project completion.

Explore our fixed, portable, and high-tech support services now!

Specifications

Brand

  • Brand
    • Honeywell RAE Systems
    • RAE

Dimension & Weight

  • Weight
    • 26 ounce av
    • 738 gram
  • Instrument Size
    • 10 in L x 3.0 in W x 2.5 in H (25.5 x 7.6 x 6.4 cm)
  • Weight range
    • 738 g gram

Category

  • Sub-Category
    • Portable Handheld VOC Monitor
    • Premium-Gas

Range

  • Humidity Range
    • 0 to 95% RH (Non-Condensing) percent
    • 0%-95% percent
  • Sensor Range
    • 0-999.9 ppm
    • 1,000 to 15,000 ppm

Temperature

  • Minimum Temperature
    • -20
    • -20 degree Celsius
    • -4 degree Fahrenheit
  • Maximum Temperature
    • 122 degree Fahrenheit
    • 50
    • 50 degree Celsius

Alarm

  • Alarm Settings
    • Additional Diagnostic Alarm and Display Message for Low Battery and Pump Stall
    • Alarm Latching with Manual Override or Automatic Reset
    • High: 3 Beeps and Flashes per sec
    • Low: 2 Beeps and Flashes per sec
  • Alarm Type
    • Audible (95 dB)
    • Visual (LED)

Features

  • Alarm Included
    • Yes
  • Series Name
    • MiniRAE 3000 + (PGM-7320)
  • Languages
    • Multi-Language Support with 10 Languages Encoded

Battery & Charging

  • Batteries Required
    • Yes
  • Batteries Included
    • Yes
  • Battery Run Time
    • 16 Hours (12 Hours with Alkaline Battery Adapter) hours

Calibration

  • Calibration Procedures
    • Reflex PID Technology Calibration Memory for 8 Calibration Gases, Alarm Limits, Span Values and Calibration Dates
  • Calibration Points
    • 2
    • 3

Certifications

  • Standards
    • Classified as Intrinsically Safe for use in Class I, Div 1, Group A, B, C, D
    • Europe: ATEX II 2G EEx ia IIC T4
    • US and Canada: CSA

Datalogging

  • Datalogging Interval
    • 1 min
  • Datalogging Capable
    • Yes
  • Datalogging
    • Standard 6 Months at 1 min Intervals
    • Standard 6 months at one-minute intervals

Display

  • Display Viewing Area
    • 28 x 43 mm
  • Display Included
    • Yes
  • Display Readout
    • Battery and Shutdown Voltage
    • Date and Time, Temperature
    • High Values
    • STEL and TWA
    • VOCs as ppm by volume (mg/m³)
  • Display Type
    • 4 Lines, LED Backlight

Features & Benefits

  • Features & Benefits
    • 3 sec Response Time
    • Highly Connectivity Capability through Multiple Wireless Module Options
    • Humidity Compensation with Built-In Humidity and Temperature Sensors
    • Large Graphic Display with Integrated Flashlight
    • Multi-Language Support with 10 Languages Encoded
    • PID Technology
    • Six-Month Datalogging
    • Third-Generation Patented PID Technology Reflex
    • VOC Detection Range from 0-15,000 ppm

Filter

  • Filter Included
    • No

Options

  • User Options
    • Alarms latching with manual override or automatic reset

Sampling

  • Sampling Pump Location
    • Horizontally and Vertically
  • Sampling Pump Included
    • Yes
  • Sampling Pump Range
    • 100 ft meter
    • 100ft (30m) meter
  • Sampling Pump Rate
    • 500

Sensor

  • Number of Sensors Included
    • 2
  • Sensor Type
    • Photoionization

Warranty

  • Warranty
    • 1 Year for Pump, Battery and Instrument
    • 3 Years for 10.6 eV Lamp

Wireless

  • Wireless Technology
    • MESH, BLE
  • Wireless Enabled
    • Yes

Standards & Certifications

  • Ingress Protection
    • IP65 Unit Running
    • IP67 Unit Off and without Flexible Probe

Additional Specifications

  • Background Gas
    • VOC
  • Combustible
    • Yes
  • Gas Type
    • VOC
  • Decibels
    • 95 decibel

Dimensions & Weight

  • Size
    • 10 in L x 3.0 in W x 2.5 in H

Technical Specifications

  • Communication and Data Download
    • Download Data and Upload Instrument Set-Up From PC Through Charging Cradle or Using BLE Module and Dedicated App
    • Wireless Data Transmission Through Built-In RF Modem
  • Wireless Approvals
    • CE RandTTE
    • FCC Part 15
  • Wireless Network
    • Mesh RAE Systems Dedicated Wireless Network
  • Wireless Range
    • BLE: Up to 15 ft (5m)
    • EchoView Host: LOS > 660 ft (200 m)
    • ProRAE Guardian and RAELink3 Mesh: LOS > 330 ft (100m)
    • ProRAE Guardian and RAEMesh Reader: LOS > 660 ft (200m)
  • Gases Detected
    • VOC
  • Keypad Type
    • 1 Operation and 2 Programming Keys, 1 Flashlight On/Off
  • Resolution
    • VOC: 0.1 ppm, 1 ppm
  • Attachment
    • Durable Bright Yellow Rubber Boot
  • Pump Draw
    • Sample from 100 ft (30m) Horizontally or Vertically
  • Sampling
    • Internal
  • EMC/RFI
    • Contact: ±4kV, Air: ±8kV, No Alarm
    • EMC Directive: 2004/108/EC
    • EMI and ESD Test: 100 MHz to 1 GHz 30V/m, No Alarm

Material Specifications

  • Housing Material
    • Rubber

Electrical Specifications

  • Power Source
    • Rechargeable, External Field-Replaceable Lithium-Ion Battery Pack, Alkaline Battery Adapter

Applications & Compatibility

  • Application
    • Civil Defense
    • Environmental and Indoor Air Quality
    • HazMat Response
    • Industrial Safety
    • Oil and Gas

Brand

  • Brand
    • Honeywell RAE Systems
    • RAE

Dimension & Weight

  • Weight
    • 26 ounce av
    • 738 gram
  • Instrument Size
    • 10 in L x 3.0 in W x 2.5 in H (25.5 x 7.6 x 6.4 cm)
  • Weight range
    • 738 g gram

Category

  • Sub-Category
    • Portable Handheld VOC Monitor
    • Premium-Gas

Range

  • Humidity Range
    • 0 to 95% RH (Non-Condensing) percent
    • 0%-95% percent
  • Sensor Range
    • 0-999.9 ppm
    • 1,000 to 15,000 ppm

Temperature

  • Minimum Temperature
    • -20
    • -20 degree Celsius
    • -4 degree Fahrenheit
  • Maximum Temperature
    • 122 degree Fahrenheit
    • 50
    • 50 degree Celsius

Alarm

  • Alarm Settings
    • Additional Diagnostic Alarm and Display Message for Low Battery and Pump Stall
    • Alarm Latching with Manual Override or Automatic Reset
    • High: 3 Beeps and Flashes per sec
    • Low: 2 Beeps and Flashes per sec
  • Alarm Type
    • Audible (95 dB)
    • Visual (LED)

Features

  • Alarm Included
    • Yes
  • Series Name
    • MiniRAE 3000 + (PGM-7320)
  • Languages
    • Multi-Language Support with 10 Languages Encoded

Battery & Charging

  • Batteries Required
    • Yes
  • Batteries Included
    • Yes
  • Battery Run Time
    • 16 Hours (12 Hours with Alkaline Battery Adapter) hours

Calibration

  • Calibration Procedures
    • Reflex PID Technology Calibration Memory for 8 Calibration Gases, Alarm Limits, Span Values and Calibration Dates
  • Calibration Points
    • 2
    • 3

Certifications

  • Standards
    • Classified as Intrinsically Safe for use in Class I, Div 1, Group A, B, C, D
    • Europe: ATEX II 2G EEx ia IIC T4
    • US and Canada: CSA

Datalogging

  • Datalogging Interval
    • 1 min
  • Datalogging Capable
    • Yes
  • Datalogging
    • Standard 6 Months at 1 min Intervals
    • Standard 6 months at one-minute intervals

Display

  • Display Viewing Area
    • 28 x 43 mm
  • Display Included
    • Yes
  • Display Readout
    • Battery and Shutdown Voltage
    • Date and Time, Temperature
    • High Values
    • STEL and TWA
    • VOCs as ppm by volume (mg/m³)
  • Display Type
    • 4 Lines, LED Backlight

Features & Benefits

  • Features & Benefits
    • 3 sec Response Time
    • Highly Connectivity Capability through Multiple Wireless Module Options
    • Humidity Compensation with Built-In Humidity and Temperature Sensors
    • Large Graphic Display with Integrated Flashlight
    • Multi-Language Support with 10 Languages Encoded
    • PID Technology
    • Six-Month Datalogging
    • Third-Generation Patented PID Technology Reflex
    • VOC Detection Range from 0-15,000 ppm

Filter

  • Filter Included
    • No

Options

  • User Options
    • Alarms latching with manual override or automatic reset

Sampling

  • Sampling Pump Location
    • Horizontally and Vertically
  • Sampling Pump Included
    • Yes
  • Sampling Pump Range
    • 100 ft meter
    • 100ft (30m) meter
  • Sampling Pump Rate
    • 500

Sensor

  • Number of Sensors Included
    • 2
  • Sensor Type
    • Photoionization

Warranty

  • Warranty
    • 1 Year for Pump, Battery and Instrument
    • 3 Years for 10.6 eV Lamp

Wireless

  • Wireless Technology
    • MESH, BLE
  • Wireless Enabled
    • Yes

Standards & Certifications

  • Ingress Protection
    • IP65 Unit Running
    • IP67 Unit Off and without Flexible Probe

Additional Specifications

  • Background Gas
    • VOC
  • Combustible
    • Yes
  • Gas Type
    • VOC
  • Decibels
    • 95 decibel

Dimensions & Weight

  • Size
    • 10 in L x 3.0 in W x 2.5 in H

Technical Specifications

  • Communication and Data Download
    • Download Data and Upload Instrument Set-Up From PC Through Charging Cradle or Using BLE Module and Dedicated App
    • Wireless Data Transmission Through Built-In RF Modem
  • Wireless Approvals
    • CE RandTTE
    • FCC Part 15
  • Wireless Network
    • Mesh RAE Systems Dedicated Wireless Network
  • Wireless Range
    • BLE: Up to 15 ft (5m)
    • EchoView Host: LOS > 660 ft (200 m)
    • ProRAE Guardian and RAELink3 Mesh: LOS > 330 ft (100m)
    • ProRAE Guardian and RAEMesh Reader: LOS > 660 ft (200m)
  • Gases Detected
    • VOC
  • Keypad Type
    • 1 Operation and 2 Programming Keys, 1 Flashlight On/Off
  • Resolution
    • VOC: 0.1 ppm, 1 ppm
  • Attachment
    • Durable Bright Yellow Rubber Boot
  • Pump Draw
    • Sample from 100 ft (30m) Horizontally or Vertically
  • Sampling
    • Internal
  • EMC/RFI
    • Contact: ±4kV, Air: ±8kV, No Alarm
    • EMC Directive: 2004/108/EC
    • EMI and ESD Test: 100 MHz to 1 GHz 30V/m, No Alarm

Material Specifications

  • Housing Material
    • Rubber

Electrical Specifications

  • Power Source
    • Rechargeable, External Field-Replaceable Lithium-Ion Battery Pack, Alkaline Battery Adapter

Applications & Compatibility

  • Application
    • Civil Defense
    • Environmental and Indoor Air Quality
    • HazMat Response
    • Industrial Safety
    • Oil and Gas
  • Brand : Honeywell RAE Systems|RAE
  • Weight : 26 ounce av|738 gram
  • Sub-Category : Portable Handheld VOC Monitor|Premium-Gas
  • Instrument Size : 10 in L x 3.0 in W x 2.5 in H (25.5 x 7.6 x 6.4 cm)
  • Weight range : 738 g gram
  • Humidity Range : 0 to 95% RH (Non-Condensing) percent|0%-95% percent
  • Sensor Range : 0-999.9 ppm|1,000 to 15,000 ppm
  • Minimum Temperature : -20|-20 degree Celsius|-4 degree Fahrenheit
  • Maximum Temperature : 122 degree Fahrenheit|50|50 degree Celsius
  • Alarm Settings : Additional Diagnostic Alarm and Display Message for Low Battery and Pump Stall|Alarm Latching with Manual Override or Automatic Reset|High: 3 Beeps and Flashes per sec|Low: 2 Beeps and Flashes per sec
  • Alarm Included : Yes
  • Alarm Type : Audible (95 dB)|Visual (LED)
  • Batteries Required : Yes
  • Batteries Included : Yes
  • Battery Run Time : 16 Hours (12 Hours with Alkaline Battery Adapter) hours
  • Calibration Procedures : Reflex PID Technology Calibration Memory for 8 Calibration Gases, Alarm Limits, Span Values and Calibration Dates
  • Calibration Points : 2|3
  • Standards : Classified as Intrinsically Safe for use in Class I, Div 1, Group A, B, C, D|Europe: ATEX II 2G EEx ia IIC T4|US and Canada: CSA
  • Datalogging Interval : 1 min
  • Datalogging Capable : Yes
  • Datalogging : Standard 6 Months at 1 min Intervals|Standard 6 months at one-minute intervals
  • Display Viewing Area : 28 x 43 mm
  • Display Included : Yes
  • Display Readout : Battery and Shutdown Voltage|Date and Time, Temperature|High Values|STEL and TWA|VOCs as ppm by volume (mg/m³)
  • Display Type : 4 Lines, LED Backlight
  • Features & Benefits : 3 sec Response Time|Highly Connectivity Capability through Multiple Wireless Module Options|Humidity Compensation with Built-In Humidity and Temperature Sensors|Large Graphic Display with Integrated Flashlight|Multi-Language Support with 10 Languages Encoded|PID Technology|Six-Month Datalogging|Third-Generation Patented PID Technology Reflex|VOC Detection Range from 0-15,000 ppm
  • Filter Included : No
  • User Options : Alarms latching with manual override or automatic reset
  • Sampling Pump Location : Horizontally and Vertically
  • Sampling Pump Included : Yes
  • Sampling Pump Range : 100 ft meter|100ft (30m) meter
  • Sampling Pump Rate : 500
  • Number of Sensors Included : 2
  • Sensor Type : Photoionization
  • Warranty : 1 Year for Pump, Battery and Instrument|3 Years for 10.6 eV Lamp
  • Wireless Technology : MESH, BLE
  • Wireless Enabled : Yes
  • Ingress Protection : IP65 Unit Running|IP67 Unit Off and without Flexible Probe
  • Background Gas : VOC
  • Combustible : Yes
  • Gas Type : VOC
  • Decibels : 95 decibel
  • Size : 10 in L x 3.0 in W x 2.5 in H
  • Communication and Data Download : Download Data and Upload Instrument Set-Up From PC Through Charging Cradle or Using BLE Module and Dedicated App|Wireless Data Transmission Through Built-In RF Modem
  • Wireless Approvals : CE RandTTE|FCC Part 15
  • Wireless Network : Mesh RAE Systems Dedicated Wireless Network
  • Housing Material : Rubber
  • Wireless Range : BLE: Up to 15 ft (5m)|EchoView Host: LOS > 660 ft (200 m)|ProRAE Guardian and RAELink3 Mesh: LOS > 330 ft (100m)|ProRAE Guardian and RAEMesh Reader: LOS > 660 ft (200m)
  • Gases Detected : VOC
  • Power Source : Rechargeable, External Field-Replaceable Lithium-Ion Battery Pack, Alkaline Battery Adapter
  • Series Name : MiniRAE 3000 + (PGM-7320)
  • Keypad Type : 1 Operation and 2 Programming Keys, 1 Flashlight On/Off
  • Application : Civil Defense|Environmental and Indoor Air Quality|HazMat Response|Industrial Safety|Oil and Gas
  • Resolution : VOC: 0.1 ppm, 1 ppm
  • Attachment : Durable Bright Yellow Rubber Boot
  • Pump Draw : Sample from 100 ft (30m) Horizontally or Vertically
  • Languages : Multi-Language Support with 10 Languages Encoded
  • Sampling : Internal
  • EMC/RFI : Contact: ±4kV, Air: ±8kV, No Alarm|EMC Directive: 2004/108/EC|EMI and ESD Test: 100 MHz to 1 GHz 30V/m, No Alarm
Data Sheet
Name
Description
File Size
Date
Size
MiniRAE 3000 + Data Sheet - FR
application/pdf 574.31 KB
7/15/2021
574.31 KB
MiniRAE 3000 + Data Sheet - ES
application/pdf 572.94 KB
7/15/2021
572.94 KB
MiniRAE 3000 + Data Sheet - PT
application/pdf 574.41 KB
7/15/2021
574.41 KB
MiniRAE 3000+ Data Sheet - DE
application/pdf 575.45 KB
7/15/2021
575.45 KB
MiniRAE 3000 + Data Sheet - RU
application/pdf 604.02 KB
7/15/2021
604.02 KB
MiniRAE 3000 + Data Sheet - IT
application/pdf 577.93 KB
7/15/2021
577.93 KB
MiniRAE 3000 + Data Sheet - NL
application/pdf 572.3 KB
7/15/2021
572.3 KB
MiniRAE 3000 + Data Sheet - EN
application/pdf 570.25 KB
7/15/2021
570.25 KB
Manuals and Guides
Name
Description
File Size
Date
Size
3000+ Series PID Quick Start Guide - ES
application/pdf 735.47 KB
7/15/2021
735.47 KB
3000+ Series PID Quick Start Guide - EN
application/pdf 737.56 KB
4/23/2024
737.56 KB
Handheld PID Monitors User Manual - EN
application/pdf 6.83 MB
12/12/2023
6.83 MB
Software
Name
Description
File Size
Date
Size
MiniRAE3000+ PGM-7320 V2.22A.35
application/zip 963.57 KB
7/15/2021
963.57 KB
Technical Note
Name
Description
File Size
Date
Size
3GPID+ Technical Note
application/pdf 436.09 KB
7/15/2021
436.09 KB
Technical Note 130 Setting Alarm Limits For Mixtures
application/pdf 788.83 KB
7/15/2021
788.83 KB
Technical Note 178 Humidity Filtering II Tube For PID Measurements In Humid Environments
application/pdf 1.09 MB
2/8/2022
1.09 MB
Technical Note 158 Conversion Of PID Readings To Methane Equivalent Or Hexane Equivalent FID Response
application/pdf 950.72 KB
7/15/2021
950.72 KB
Technical Note 118 Using RAE Systems PIDs For Measuring Soil Headspace And In Soil Vapor Extractions
application/pdf 813.81 KB
2/8/2022
813.81 KB
Technical Note 132 Upgrading Firmware On RAE Systems Instruments
application/pdf 820.7 KB
9/14/2022
820.7 KB
Technical Note 120 Measuring PID Correction Factors For Volatile Compounds With RAE Systems Instruments
application/pdf 624.75 KB
7/15/2021
624.75 KB
Technical Note 141 Monitor Remote Control Functions
application/pdf 984.3 KB
8/3/2022
984.3 KB
Technical Note 140 The Effect Of Extension Tubing Volume And Vacuum On Sample Flow For RAE Systems Instruments
application/pdf 911.54 KB
7/15/2021
911.54 KB
Technical Note 174Decontamination Procedures For Monitors Exposed To Chemical Warfare Agents
application/pdf 814.89 KB
7/15/2021
814.89 KB
Technical Note 170 USB To Serial Port Adapter
application/pdf 1.26 MB
7/15/2021
1.26 MB
Lithium-Ion Batteries Technical Note 166
application/pdf 926.7 KB
7/15/2021
926.7 KB
Technical Note 122 Compliance of RAE Systems PIDs with EPA Method
application/pdf 790.79 KB
7/15/2021
790.79 KB
Technical Note 165 Combating Drift In Portable And Fixed PIDs
application/pdf 930.69 KB
7/15/2021
930.69 KB
A Guideline for Pid Instrument Response
application/pdf 1.45 MB
8/3/2022
1.45 MB
Technical Note 148 Defining The Interval Between Calibration Checks
application/pdf 839.32 KB
12/14/2022
839.32 KB
Certificate
Name
Description
File Size
Date
Size
EU Declaration of Conformity - MiniRAE 3000+ and MiniRAE Lite+
application/pdf 203.41 KB
11/24/2021
203.41 KB
EU Declaration of Conformity - MiniRAE 3000+ and MiniRAE Lite+
application/pdf 240.01 KB
2/27/2023
240.01 KB
Application Note
Name
Description
File Size
Date
Size
Application Note 221: Using PIDs To Assess Exposure Risk In Unknown Environments
application/pdf 964.88 KB
7/15/2021
964.88 KB
Application Note 000: RAE Systems PID Overview
application/pdf 1.1 MB
7/15/2021
1.1 MB
Application Note 214: Environmental Applications For PIDs
application/pdf 836.36 KB
7/15/2021
836.36 KB
Application Note 201: Measuring Ammonia (NH3) With Photoionization Detectors
application/pdf 655.59 KB
7/15/2021
655.59 KB
Application Note 216: Using PIDs In Terrorist Chemical Attacks
application/pdf 1.07 MB
7/15/2021
1.07 MB
Application Note 203: PIDs As HazMat Response Tools
application/pdf 1.26 MB
7/15/2021
1.26 MB
Name
Description
File Size
Date
Size
MiniRAE3000+ PGM-7320 V2.22A.35
application/zip 963.57 KB
7/15/2021
963.57 KB
3GPID+ Technical Note
application/pdf 436.09 KB
7/15/2021
436.09 KB
Technical Note 130 Setting Alarm Limits For Mixtures
application/pdf 788.83 KB
7/15/2021
788.83 KB
Technical Note 178 Humidity Filtering II Tube For PID Measurements In Humid Environments
application/pdf 1.09 MB
2/8/2022
1.09 MB
Technical Note 158 Conversion Of PID Readings To Methane Equivalent Or Hexane Equivalent FID Response
application/pdf 950.72 KB
7/15/2021
950.72 KB
Application Note 221: Using PIDs To Assess Exposure Risk In Unknown Environments
application/pdf 964.88 KB
7/15/2021
964.88 KB
Technical Note 118 Using RAE Systems PIDs For Measuring Soil Headspace And In Soil Vapor Extractions
application/pdf 813.81 KB
2/8/2022
813.81 KB
3000+ Series PID Quick Start Guide - ES
application/pdf 735.47 KB
7/15/2021
735.47 KB
MiniRAE 3000 + Data Sheet - FR
application/pdf 574.31 KB
7/15/2021
574.31 KB
EU Declaration of Conformity - MiniRAE 3000+ and MiniRAE Lite+
application/pdf 203.41 KB
11/24/2021
203.41 KB
Application Note 000: RAE Systems PID Overview
application/pdf 1.1 MB
7/15/2021
1.1 MB
Technical Note 132 Upgrading Firmware On RAE Systems Instruments
application/pdf 820.7 KB
9/14/2022
820.7 KB
EU Declaration of Conformity - MiniRAE 3000+ and MiniRAE Lite+
application/pdf 240.01 KB
2/27/2023
240.01 KB
Technical Note 120 Measuring PID Correction Factors For Volatile Compounds With RAE Systems Instruments
application/pdf 624.75 KB
7/15/2021
624.75 KB
Application Note 214: Environmental Applications For PIDs
application/pdf 836.36 KB
7/15/2021
836.36 KB
MiniRAE 3000 + Data Sheet - ES
application/pdf 572.94 KB
7/15/2021
572.94 KB
Technical Note 141 Monitor Remote Control Functions
application/pdf 984.3 KB
8/3/2022
984.3 KB
MiniRAE 3000 + Data Sheet - PT
application/pdf 574.41 KB
7/15/2021
574.41 KB
Application Note 201: Measuring Ammonia (NH3) With Photoionization Detectors
application/pdf 655.59 KB
7/15/2021
655.59 KB
Application Note 216: Using PIDs In Terrorist Chemical Attacks
application/pdf 1.07 MB
7/15/2021
1.07 MB
MiniRAE 3000+ Data Sheet - DE
application/pdf 575.45 KB
7/15/2021
575.45 KB
3000+ Series PID Quick Start Guide - EN
application/pdf 737.56 KB
4/23/2024
737.56 KB
Technical Note 140 The Effect Of Extension Tubing Volume And Vacuum On Sample Flow For RAE Systems Instruments
application/pdf 911.54 KB
7/15/2021
911.54 KB
MiniRAE 3000 + Data Sheet - RU
application/pdf 604.02 KB
7/15/2021
604.02 KB
MiniRAE 3000 + Data Sheet - IT
application/pdf 577.93 KB
7/15/2021
577.93 KB
Technical Note 174Decontamination Procedures For Monitors Exposed To Chemical Warfare Agents
application/pdf 814.89 KB
7/15/2021
814.89 KB
Technical Note 170 USB To Serial Port Adapter
application/pdf 1.26 MB
7/15/2021
1.26 MB
Handheld PID Monitors User Manual - EN
application/pdf 6.83 MB
12/12/2023
6.83 MB
Lithium-Ion Batteries Technical Note 166
application/pdf 926.7 KB
7/15/2021
926.7 KB
Technical Note 122 Compliance of RAE Systems PIDs with EPA Method
application/pdf 790.79 KB
7/15/2021
790.79 KB
Application Note 203: PIDs As HazMat Response Tools
application/pdf 1.26 MB
7/15/2021
1.26 MB
MiniRAE 3000 + Data Sheet - NL
application/pdf 572.3 KB
7/15/2021
572.3 KB
Technical Note 165 Combating Drift In Portable And Fixed PIDs
application/pdf 930.69 KB
7/15/2021
930.69 KB
A Guideline for Pid Instrument Response
application/pdf 1.45 MB
8/3/2022
1.45 MB
Technical Note 148 Defining The Interval Between Calibration Checks
application/pdf 839.32 KB
12/14/2022
839.32 KB
MiniRAE 3000 + Data Sheet - EN
application/pdf 570.25 KB
7/15/2021
570.25 KB
Name
Date
Size
Showing 1 to 1 of 1 parts
Product Number
Part Number
Description
Leaf Category
Thumbnail
059-9120-100
Hard Case Kit, MiniRAE3000+, universal accessories with gas and flow regulator
https://s7d1.scene7.com/is/image/Honeywell65/MiniRAE3000
059-9120-100
Hard Case Kit, MiniRAE3000+, universal accessories with gas and flow regulator
https://s7d1.scene7.com/is/image/Honeywell65/MiniRAE3000
Feedback